Model Number 3788
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bacterial Infection (1735); Post Operative Wound Infection (2446)
Event Date 03/07/2014
Event Type  Injury  
Event Description
It was reported the patient experienced fluid in his scs ipg pocket site.Subsequently, the physician aspirated the fluid and prescribed antibiotics.It was also reported the patient later went to the emergency room due to a fever.As a result, the physician decided to explant the scs system.The patient felt "better" following the surgical intervention.F/u identified cultures dowed positive for staph aureus.Additionally, a user facility report was rec'd by the manufacturer.
 
Manufacturer Narrative
Method: the device history and sterilization records were reviewed.Results: the device history and sterilization records reviewed were found to meet specifications and no anomalies related to this event were found.Conclusion: the cause of the reported complaint could not be determined from the review of the dhr and sterilization records.Sjm has limited info related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL, NEUROMODULATION
plano TX
Manufacturer Contact
rachel blackwell
6901 preston rd.
plano, TX 75024
9725264637
MDR Report Key3764087
MDR Text Key15806919
Report Number1627487-2014-03211
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 03/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/01/2015
Device Model Number3788
Device Lot Number4387100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/10/2014
Initial Date FDA Received04/01/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS LEAD, MODEL 3186 (2); SCS ANCHOR, MODEL 1192 (2); IMPLANT DATE:
Patient Outcome(s) Hospitalization;
Patient Age57 YR
Patient Weight99
-
-