• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW WOUND MANAGEMENT RENASYS GO NEGATIVE PRESSURE WOUND THERAPY; PUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW WOUND MANAGEMENT RENASYS GO NEGATIVE PRESSURE WOUND THERAPY; PUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED) Back to Search Results
Model Number 66800164
Device Problems Air Leak (1008); Device Alarm System (1012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/07/2012
Event Type  malfunction  
Event Description
Failure to alarm: no error message in display but dressing air leaking and no alarm went off.
 
Manufacturer Narrative
This medical device report (mdr) is being submitted by smith & nephew as a result of a retrospective review of complaints, including adverse events, as part of corrective and remediation actions following the issuance of the 483 issued november 26, 2013 to smith & nephew, inc.(b)(4).Smith & nephew is performing a two-year retrospective review of complaint files to re-assess reportability criteria and reporting decisions made for complaints recorded during the time period under review.This complaint has been re-assessed in accordance under the provisions of 21cfr 803.50 and deemed reportable as an mdr.We are submitting one (1) initial, 30 day report, medwatch fda form 3500a, for a purported device malfunction which occurred while using the renasys go negative pressure wound therapy.The issue was not confirmed.The root cause of the reported issue cannot be determined.It is recommended that the device be returned for evaluation.However, as evidenced by the serial number this pump is more than 28 months old.At this time the internal pump reach the end of the life cycle.Smith & nephew has issued (b)(4) to initiate renasys product labeling updates.These labelling updates will provide enhanced information on the functionality of the alarms as they relate to the design of the device, and scenarios that can occur in clinical settings which may impact alarm functionality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS GO NEGATIVE PRESSURE WOUND THERAPY
Type of Device
PUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)
Manufacturer (Section D)
SMITH & NEPHEW WOUND MANAGEMENT
970 lake carillon drive
suite 110
st petersburg FL 33716
Manufacturer (Section G)
SMITH & NEPHEW WOUND MANAGEMENT
970 lake carillon drive
suite 110
st petersburg FL 33716
Manufacturer Contact
terry mcmahon
970 lake carillon drive
suite 110
st petersburg, FL 33716
7273993785
MDR Report Key3764188
MDR Text Key4468281
Report Number3006760724-2014-00242
Device Sequence Number1
Product Code BTA
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K083375
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Model Number66800164
Was Device Available for Evaluation? No
Distributor Facility Aware Date05/21/2012
Event Location Hospital
Date Manufacturer Received05/21/2012
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
Patient Outcome(s) Other;
-
-