• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW WOUND MANAGEMENT RENASYS GO NEGATIVE PRESSURE WOUND THERAPY DEVICE; PUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW WOUND MANAGEMENT RENASYS GO NEGATIVE PRESSURE WOUND THERAPY DEVICE; PUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED) Back to Search Results
Model Number 66800164
Device Problems Device Alarm System (1012); Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/12/2011
Event Type  malfunction  
Event Description
Failure to alarm: alarm doesn´t go off although there´s a leakage.
 
Manufacturer Narrative
This medical device report (mdr) is being submitted by smith & nephew as a result of a retrospective review of complaints, including adverse events, as part of corrective and remediation actions following the issuance of the 483 issued november 26, 2013 to smith & nephew, inc.Fei no: 3006160724.Smith & nephew is performing a two-year retrospective review of complaint files to re-assess reportability criteria and reporting decisions made for complaints recorded during the time period under review.This complaint has been re-assessed in accordance under the provisions of 21cfr 803.50 and deemed reportable as an mdr.We are submitting one (1) initial, 30 day report, medwatch fda form 3500a, for a purported device malfunction which occurred while using the renasys go negative pressure wound therapy device.Smith & nephew received the above named product reported as a field complaint for failure to alarm.This issue could not be confirmed as the device was not received for evaluation.The probable root cause is most likely due to circumstances unrelated to a malfunctioning of the device, possibly related to a kit issue or dressing method used.The device history for this device was reviewed and no issues were revealed.Neither a definitive root cause nor a corrective action could be determined without evaluation of the device.Smith & nephew has issued (b)(4) to initiate renasys product labeling updates.These labelling updates will provide enhanced information on the functionality of the alarms as they relate to the design of the device, and scenarios that can occur in clinical settings which may impact alarm functionality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS GO NEGATIVE PRESSURE WOUND THERAPY DEVICE
Type of Device
PUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)
Manufacturer (Section D)
SMITH & NEPHEW WOUND MANAGEMENT
970 lake carillon drive
suite 110
st petersburg FL 33716
Manufacturer (Section G)
SMITH AND NEPHEW WOUND MANAGEMENT
970 lake carillon drive
suite 110
st. petersburg FL 33716
Manufacturer Contact
terry mcmahon
970 lake carillon drive
suite 110
st petersburg, FL 33716
7273993785
MDR Report Key3764786
MDR Text Key16800721
Report Number3006760724-2014-00257
Device Sequence Number1
Product Code BTA
Combination Product (y/n)N
PMA/PMN Number
K083375
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number66800164
Was Device Available for Evaluation? No
Distributor Facility Aware Date05/12/2011
Event Location Hospital
Date Manufacturer Received05/12/2011
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
Patient Outcome(s) Other;
-
-