• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE FLORIDA OPERATIONS AC-POWERED ADJUSTABLE HOSPITAL BED, INCLUDING RAILS; 880.5100

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE FLORIDA OPERATIONS AC-POWERED ADJUSTABLE HOSPITAL BED, INCLUDING RAILS; 880.5100 Back to Search Results
Model Number 5310IVC
Device Problem Torn Material (3024)
Patient Problem No Information (3190)
Event Date 03/20/2014
Event Type  malfunction  
Event Description
Dealer advised pendant cord is severed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC-POWERED ADJUSTABLE HOSPITAL BED, INCLUDING RAILS
Type of Device
880.5100
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
gregory stevens
one invacare way
elyria, OH 44035
8003336900
MDR Report Key3771799
MDR Text Key4374653
Report Number1031452-2014-01069
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 03/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number5310IVC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/24/2014
Initial Date FDA Received04/24/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-