|
Model Number 72201108 |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 03/28/2014 |
Event Type
malfunction
|
Event Description
|
During a labrum repair in the shoulder, it was reported that the tip of the drill broke at the hub inside the patient¿s bone.After a 2.3mm anchor was inserted, it was noticed that the tip was still inside the drilled hole within bone.An x-ray was done to confirm the piece was present.A 2.9mm anchor was also used without further incident.The patient¿s post-op condition was reported as fine.Clarification from the sales rep.States that an mri was performed several days post-op and the broken tip of the drill was not in the bone but in the soft tissue of the patient¿s right shoulder.The patient will be rescheduled for additional procedure to remove the broken tip.The customer has retained the drill; however, a picture of the device was taken and will be forwarded to be included in the complaint file.
|
|
Manufacturer Narrative
|
The customer has retained the device therefore, no evaluation is possible.(b)(4).
|
|
Manufacturer Narrative
|
A device was not returned for evaluation.However the reporter was able to provide a picture of the complaint device.It was reviewed and confirmed to be broken, as the cutting edges were not attached to the shaft of the device.However due the quality of the picture it is difficult to determine the exact nature and location of the break.A review of the manufacturing records identified that the device was manufactured april 2008, and no issues were identified during manufacture.No further analysis could be performed therefore a conclusion regarding root cause could not be made.(b)(4).
|
|
Search Alerts/Recalls
|
|
|