• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION, INC AUTOPULSE® RESUSCITATION SYSTEM MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION, INC AUTOPULSE® RESUSCITATION SYSTEM MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/20/2014
Event Type  malfunction  
Event Description
Customer received a call regarding a (b)(6) male, weighing (b)(6) pounds.The arrest was witnessed by a bystander.The autopulse platform performed compressions for approximately 2 minutes and then the autopulse nimh battery stopped working.The battery was replaced.However, the platform indicated that this second battery was also "dead" (depleted).No adverse patient sequelae was reported.Patient regained a pulse and was transported to the hospital.
 
Manufacturer Narrative
Customer reported that the batteries were checked and rotated on the morning of the reported event.Product in complaint will not be returned.Therefore, physical investigation cannot be performed.A supplemental report will be filed if the product in complaint is returned and investigation has been completed.Please see the following related mfr.Reports: #3003793491-2014-00217 for autopulse nimh battery with sn: (b)(4); #3003793491-2014-00218 for autopulse nimh battery with sn: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION SYSTEM MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION, INC
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION, INC
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave.
san jose, CA 95131
4084192957
MDR Report Key3776850
MDR Text Key4440419
Report Number3003793491-2014-00219
Device Sequence Number1
Product Code DRM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Paramedic
Type of Report Initial
Report Date 04/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/03/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age68 YR
Patient Weight109
-
-