MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO CAPITAL TEMP PUMP PROFESSIONAL; PACK, HOT OR COLD, WATER CIRCULATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO CAPITAL TEMP PUMP PROFESSIONAL; PACK, HOT OR COLD, WATER CIRCULATING Back to Search Results
Catalog Number TP700
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Burn(s) (1757); Swelling (2091)
Event Date 04/03/2014
Event Type  Injury  
Event Description
It was reported by repair work order that the patient received several blisters and burn marks during use of the device.
 
Event Description
It was reported by repair work order that the patient received several blisters and burn marks during use of the device.
 
Manufacturer Narrative
After investigation, the alleged skin damage event was unable to be confirmed and no specific product malfunction was identified that may have caused the issue.After speaking with the customer, evaluating the returned product details, and reviewing the product manual, it was identified that the cause as to why the patient may have had the skin damage event was likely from operator error and leaving the patient exposed to too much compression and shearing forces which can lead to skin damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPITAL TEMP PUMP PROFESSIONAL
Type of Device
PACK, HOT OR COLD, WATER CIRCULATING
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
anthony noora
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key3778015
MDR Text Key4373338
Report Number0001831750-2014-02938
Device Sequence Number1
Product Code ILO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberTP700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/03/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-