During a rotator cuff repair utilizing the truepass needle, single pack, sterile it was reported that the tip broke off the device.It was not retrieved.The passer was the self-retrieving design.The broken piece wasn¿t found ¿ it may have been sucked up by the shaver; however, this cannot be confirmed.The patient¿s post-operative condition was fine.
|
The two devices were investigated together.Both needles were dimensionally measured in thickness and width and confirmed to be within specification.The weld is intact and the needles demonstrate evidence of duraglide over the required length.One device has damage along the needle indicative of contact with a metal or hard surface.This device was exposed to loads not expected with passage of suture through soft tissue.The needle was passed without visualization (blind) and a secondary grasping instrument was also used blindly to grab the suture before retraction of the needle.This allowed the grasper to grip the needle directly and cause the damage noted above.Per ifu ¿prior to use, inspect the device to ensure it is not damaged.Do not use a damaged device.As with any surgical device, careful attention should be exercised to ensure that excessive force is not placed on this device.Excessive force can result in the failure of this device.Ensure adequate visibility when using the device.(b)(4).
|