Device 1 of 3.Reference mfr report #: 1627487-2014-05270.Reference mfr report #: 1627487-2014-05271.The pt had two, 3146 leads from the same lot.It was reported the pt lost stimulation from the 3146 leads after falling.The 3286 lead provided inadequate coverage/ineffective therapy and stimulation in an unintended area, i.E., abdomen.Reprogramming to provide resolution was unsuccessful.X-rays were taken and revealed no anomalies.As a result, the pt's lead were explanted and replaced (with a 3186 lead and a 3286 lead).During the procedure, the header of the ipg was found to be damaged.In turn, the ipg was also explanted and replaced.Effective therapy was restored post-op.
|