Initially, it was reported that the physician was experiencing intermittent communication issues with the handheld.Troubleshooting performed by a company representative did not identify any communication problems.The physician requested a new handheld and the handheld was returned for analysis.Analysis of the handheld was completed on (b)(6) 2014.During the analysis it was identified that the handheld was unable to establish communication with a known good wand and generator.The cause for the anomaly is associated with a broken wire connection in the serial cable.Once the wire was soldered onto the serial cable pcb, no further anomalies were identified.
|