• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO DUAL SPEED CEMENT INJECTION GUN; MIXER, CEMENT, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO DUAL SPEED CEMENT INJECTION GUN; MIXER, CEMENT, FOR CLINICAL USE Back to Search Results
Catalog Number 0206600000
Device Problem Metal Shedding Debris (1804)
Patient Problem No Patient Involvement (2645)
Event Date 04/11/2014
Event Type  malfunction  
Event Description
During inspection at the user facility, the dual speed cement injection gun was producing metal shavings.It was reported that there was no associated procedure.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
 
Manufacturer Narrative
Based on the manufacturing engineer¿s evaluation of the device, the cause was determined to be the rod catching on the bushings due to interference.The device was repaired and returned to the user facility.
 
Event Description
During inspection at the user facility the dual speed cement injection gun was producing metal shavings.It was reported that there was no associated procedure.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
 
Manufacturer Narrative
A follow up report will be filed after the quality investigation has been completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL SPEED CEMENT INJECTION GUN
Type of Device
MIXER, CEMENT, FOR CLINICAL USE
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3789625
MDR Text Key4537905
Report Number0001811755-2014-01602
Device Sequence Number1
Product Code JDZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0206600000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received07/18/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-