• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE 2 ROVER ULTRA (120V); APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE 2 ROVER ULTRA (120V); APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Catalog Number 0702001000
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 04/09/2014
Event Type  malfunction  
Event Description
It was reported that the neptune 2 rover ultra (120v) had bare wires showing through the insulation of the power cord during repair by a stryker field service technician.There was no patient involvement and no adverse consequences associated with the device.
 
Event Description
It was reported that the neptune 2 rover ultra (120v) had bare wires showing through the insulation of the power cord during repair by a stryker field service technician.There was no patient involvement and no adverse consequences associated with the device.
 
Manufacturer Narrative
The field service technician was able to confirm the reported event.He found that the power cord had exposed wires.The root cause was determined to be due to the power cord.The power cord¿s insulation can break down exposing the bare copper wires due to being exposed to rough surfaces, sharp edges, or corners.The device was repaired at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPTUNE 2 ROVER ULTRA (120V)
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3790606
MDR Text Key4434961
Report Number0001811755-2014-01612
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132671
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0702001000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/27/2014
Initial Date FDA Received05/05/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/29/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/14/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-