MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO ASEPTIC BATTERY HOUSING; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO ASEPTIC BATTERY HOUSING; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 4126120000
Device Problems Disassembly (1168); Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 04/01/2014
Event Type  malfunction  
Event Description
It was reported that the aseptic housing fractured and detached from the top housing to the aseptic bottom housing where battery is placed during inspection at user facility.No associated procedure, no patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
 
Manufacturer Narrative
The reported event, housings disassembled, could not be confirmed as the battery housings were not available for evaluation.Based on a review of previous cases with a similar reported event and risk documentation, the most likely cause of the failure could be due to improper insertion of the batteries into the battery housings or they were damaged due to impact.Device is scrapped by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASEPTIC BATTERY HOUSING
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3799002
MDR Text Key4479820
Report Number0001811755-2014-01655
Device Sequence Number1
Product Code KIJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K972367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4126120000
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/11/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-