STRYKER INSTRUMENTS-KALAMAZOO ASEPTIC BATTERY HOUSING; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 4126120000 |
Device Problems
Disassembly (1168); Battery Problem (2885)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/01/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that the aseptic housing fractured and detached from the top housing to the aseptic bottom housing where battery is placed during inspection at user facility.No associated procedure, no patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
The reported event, housings disassembled, could not be confirmed as the battery housings were not available for evaluation.Based on a review of previous cases with a similar reported event and risk documentation, the most likely cause of the failure could be due to improper insertion of the batteries into the battery housings or they were damaged due to impact.Device is scrapped by the customer.
|
|