The customer reported discrepant beta human chorionic gonadotrophin (bhcg) result, for one patient, involving the access® total bhcg assay used in conjunction with the unicel® dxi 600 access® immunoassay system.The original result was above the normal reference range and was not released from the laboratory.There was no patient impact associated with this event.Subsequent testing of the patient's sample, on the same instrument, recovered lower bhcg results, within the normal reference range.Per laboratory protocol, bhcg results that are less than (<) 5 miu/ml are interpreted as negative.And patient samples with results less than (<) 50 miu/ml are reanalyzed.The customer stated the serum sample was clear in appearance and 24 hours old, following collection.No sample integrity issues were noted.The customer indicated quality control (qc) and system check were within specifications prior to and at the time of the event.
|