The blender oxygen output was higher than the setting on the unit.These findings would not explain patient's change of condition.Setting of 21% - read at 74.5%, 60% setting read 85.5%, 90% setting read 98.5%) air pump output set at 30 psi - actual 31.8 psi.O2 blender removed from use.These findings would not explain patient's change of condition.Patient was placed on the ttav equipment, but was desaturating.Connections checked and no leak was detected, patient was receiving adequate flow.The patient continued to desaturate and was placed on wall oxygen at 5 liter with immediate improvement in saturation.The equipment removed from the room, marked defective and sequestered for evaluation of the o2 blender and circuit.
|