• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE Back to Search Results
Catalog Number UNKNOWN
Device Problems Difficult to Insert (1316); Peeled/Delaminated (1454); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/18/2014
Event Type  malfunction  
Event Description
It was reported that the balance middle weight (bmw) guide wire was used but could not be inserted into the non-abbott introducer.The guide wire and introducer were removed as a single unit.After removal out of the patient anatomy, it was noted that the coating was separated and moved forward remaining on the core but creating a "step" in the guide wire shaft.Another bmw guide wire was used to successfully conclude the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).It is indicated that the device is not returning for evaluation; therefore a failure analysis of the complaint device could not be completed.The lot history record for this product could not be reviewed and a similar incident query could not be performed because the product was not returned for evaluation and the part and lot numbers were not reported.Based on the reviewed information, no product deficiency was identified.Event date: date has been estimated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE
Type of Device
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key3812867
MDR Text Key17386686
Report Number2024168-2014-03066
Device Sequence Number1
Product Code DQX
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K013833
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 04/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/22/2014
Initial Date FDA Received05/15/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-