• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL CDD-SOLIS AMBULATORY INFUSION PUMP; MEA - PUMP, INFUSION, PCA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL CDD-SOLIS AMBULATORY INFUSION PUMP; MEA - PUMP, INFUSION, PCA Back to Search Results
Model Number 21-2110
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
A report was received that stated the pump alarmed "error code 44510".No pt injury or adverse event was reported.
 
Manufacturer Narrative
The suspect device was received for eval.Review of the device history record and testing of the device confirmed.Error code-44510.Therefore, the circuit board was replaced.Following repair, the device passed all function and delivery testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDD-SOLIS AMBULATORY INFUSION PUMP
Type of Device
MEA - PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL
st. paul MN
Manufacturer Contact
pete hirte
1265 grey fox rd.
st. paul, MN 55112
6516287384
MDR Report Key3814996
MDR Text Key4390839
Report Number2183502-2014-00248
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072144
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-2110
Device Catalogue Number21-2111-0100-51
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/21/2014
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/23/2014
Device Age2 MO
Event Location Hospital
Initial Date Manufacturer Received 03/26/2014
Initial Date FDA Received04/24/2014
Date Device Manufactured01/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-