Catalog Number 810081L
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/28/2014
Event Type  malfunction  
Event Description
It was reported that a patient underwent an obturator sling procedure on (b)(6) 2014.During placement of the mesh, the conical tip fractured.This occured during passage of the helical passer.The device was backed out of the patient.The tip was adjusted and then passed successfully.The procedure was completed with same device on second pass.There were no adverse patient consequences.
 
Manufacturer Narrative
(b)(4).Conclusion: the product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Manufacturer Narrative
Date sent to the fda: 06/12/2014, it was reported that the device came in contact with the bone during the procedure.No additional information was provided.Conclusion: the actual device involved in this event was returned for evaluation.The tip of the plastic tube is bent.The location of the bend on the needle indicates it could not have happened if the needle was properly locked on the guide as it is when the product is packaged at the manufacturing site.This suggests that the needle was removed from the bulge that holds the needle at the guide.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT OBTURATOR SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20 2000 neuchatel
neuchatel
SZ  
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key3817684
MDR Text Key4500475
Report Number2210968-2014-06293
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K033568
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2014
Device Catalogue Number810081L
Device Lot Number3744205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2014
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/27/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-