• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Failure to Interrogate (1332); Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/28/2014
Event Type  malfunction  
Event Description
The handheld device and associated flashcard were returned for analysis where no anomalies or device issues were found.The serial cable was not returned therefore an evaluation could not be performed on that component.
 
Event Description
It was reported that the physician was unable to interrogate multiple patients with the programming system.Troubleshooting performed by a company representative narrowed the problem to the handheld serial cable.It was reported that when the cable was manipulated into a certain position the device would interrogate.The physician opted for a new programming tablet rather than a new serial cable given the age of the handheld.The handheld and serial cable are expected to be returned for analysis, but have not been received to date.
 
Manufacturer Narrative
Device manufacturing records were reviewed.Review of manufacturing records of the handheld confirmed all quality tests were passed prior to distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3824632
MDR Text Key5244215
Report Number1644487-2014-01343
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number874814
Other Device ID NumberVERSION 8.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 12/02/2014
Initial Date FDA Received05/21/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/16/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/22/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-