MAUDE Adverse Event Report: HILL-ROM, INC. CLINITRON RITEHITE BED; BED, AIR FLUIDIZED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM, INC. CLINITRON RITEHITE BED; BED, AIR FLUIDIZED Back to Search Results
Model Number 0800
Device Problem Device Slipped (1584)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/27/2014
Event Type  malfunction  
Event Description
The account reported the brakes do not lock.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Manufacturer Narrative
The technician found the brakes were not fully engaged, user error.Per the hill-rom user manual, warning: pts may use the bed for support while entering or exiting; if the unit moves unexpectedly, pt injury could occur.When the unit is unattended, ensure that both brakes are locked.The brakes for the clinitron bed are located at the right, head end and the left, foot end of the unit.To apply the brakes, step on the lower end of the brake lever to lock the wheels.To release the brakes, apply inward pressure to the upper end of the brake lever.The technician in-serviced the account on the brake functions to resolve the issue.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITRON RITEHITE BED
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
jennifer morris
1069 state route 46 east
batesville, IN 47006
8129313121
MDR Report Key3842624
MDR Text Key4680219
Report Number1824206-2014-00577
Device Sequence Number1
Product Code INX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Unknown
Type of Report Initial
Report Date 01/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/27/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-