Catalog Number ST0A19P
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/14/2014
Event Type  malfunction  
Event Description
It was reported during surgery doctor implanted smart toe, metal did not expand in canal within a reasonable time frame.Doctor then implanted larger size which led to a delay in surgery.
 
Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Manufacturer Narrative
Evaluation summary: the reported event that smart toe implant had no expansion during surgery could not be confirmed since the returned device is conforming to specification.Therefore, the case could be classified as a non-issue.Based on investigation, the root cause of the determined no expansion problem was attributed to a user related issue.The smart toe implant expansion problem was caused by mismanagement of the temperature.When the blood is arrested for a long period, the smart toe can take longer to expand because the temperature of the body is too low.Using warm sterile solution between 37°c and 40°c to bath phalanges allows the smarttoe implant to expand properly.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.Indications for any material, manufacturing or design related problems were not determined in the investigation.If any further information is provided, the investigation report will be updated.
 
Event Description
It was reported during surgery doctor implanted smart toe, metal did not expand in canal within a reasonable time frame.Doctor then implanted larger size which led to a delay in surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAMEDULLARY ARTHRODESIS IMPLANT SMART TOE II 19MM / 10° ANGLE FOR PIP ARTHRO
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
STRYKER TRAUMA SELZACH
bohnackerweg 1
postfach
selzach 2545
CH  2545
Manufacturer (Section G)
STRYKER TRAUMA SELZACH
bohnackerweg 1
postfach
selzach 2545
CH   2545
Manufacturer Contact
cécile lefeuvre
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key3844594
MDR Text Key4685496
Report Number0008031020-2014-00267
Device Sequence Number1
Product Code HTY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112197
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Catalogue NumberST0A19P
Device Lot NumberF004869PCBQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2014
Initial Date FDA Received06/03/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/29/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/09/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age52 YR
-
-