HILL-ROM, INC. CLINITRON RITEHITE BED; BED, AIR FLUIDIZED Back to Search Results
Model Number 0800
Device Problem Device Slipped (1584)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/12/2014
Event Type  malfunction  
Event Description
The technician reported the brake will not keep the wheel from rolling while in the brake mode.The bed was located in the hill-rom service center.There was no patient/user injury reported.(b)(4).
 
Manufacturer Narrative
The technician found during a between patient inspection that the brake caster would not hold due to the rubber on the brake pad was worn off most likely due to normal wear and tear.The technician replaced the brake caster to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITRON RITEHITE BED
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
jennifer morris
1069 state route 46 east
batesville, IN 47006
8129313121
MDR Report Key3844827
MDR Text Key4661252
Report Number1824206-2014-00827
Device Sequence Number1
Product Code INX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/12/2014
Initial Date FDA Received03/12/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/1997
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-