The investigation determined that a non-reproducible, higher than expected glucose result was obtained from a patient sample while using the vitros 5600 integrated system.There was no indication that an instrument issue contributed to the event.The investigation determined the sample was not processed in accordance with the tube manufacturer¿s recommendations.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.The investigation could not determine a definitive root cause.However, a pre-analytical sample handling issue and a reagent issue cannot be ruled out as potential contributing factors.
|