ST. JUDE MEDICAL - NEUROMODULATION EON MINI; SCS IPG Back to Search Results
Model Number 3788
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 03/28/2013
Event Type  Injury  
Event Description
It was reported the patient experienced pocket heating while recharging.A replacement charger was sent.The replacement charger resolved the issue.On 08/01/2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.Reference mfr report #1627487-2014-01325.
 
Manufacturer Narrative
This ipg serial number was included in a field advisory and a field correction.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
melissa nkematu
6901 preston rd.
plano, TX 75024
9723092520
MDR Report Key3848317
MDR Text Key21735296
Report Number1627487-2014-01324
Device Sequence Number1
Product Code GZN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Remedial Action Other
Type of Report Initial
Report Date 04/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2013
Device Model Number3788
Device Lot Number3404942
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-002-R
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS LEAD: MODEL 3186 (2)
Patient Outcome(s) Other;
Patient Age58 YR
-
-