• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO PHILIPPINES CORPORATION SURGUARD2 SAFETY NEEDLE; TERUMO NEOLUS NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO PHILIPPINES CORPORATION SURGUARD2 SAFETY NEEDLE; TERUMO NEOLUS NEEDLE Back to Search Results
Catalog Number SG2-2038
Device Problem Use of Device Problem (1670)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
The user facility reported device broke during administering of vivitrol suspension.Follow-up communication with the user facility confirmed; (1) patient may have received only 0.4ml of the medication because only 3ml was left in the syringe; and (2) there was no injury as a result.
 
Manufacturer Narrative
The involved device was not returned for evaluation, which limits the failure investigation.The investigation was based upon assessment of user facility information and evaluation of reserve samples from the reported lot and current lot samples.No abnormalities or defects were noted on visual inspection and all samples passed functional testing.A simulation test was performed on a retention sample.The safety needle was securely attached to a 5ml syringe.Sterile water for injection was withdrawn from the vial and then injected into a simulated arm at a 90 degree angle as well as a 45 degree angle.No break on the needle and hub was noted.A review of the device history records and inspection records indicated that there were no production related problems.A review of the complaint files confirmed that this lot number has not been reported previously.Although the cause for the reported break cannot be definitively determined based upon the available information, there is no indication that the event was related to a pre-existing device defect.All available information has been placed on file in quality assurance for appropriate tracking, trending and follow-up.(b)(4).
 
Manufacturer Narrative
(b)(4).
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGUARD2 SAFETY NEEDLE
Type of Device
TERUMO NEOLUS NEEDLE
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna MD 21921
RP  21921
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna MD 21921
RP   21921
Manufacturer Contact
cathleen hargreaves
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key3848917
MDR Text Key18728354
Report Number3003902955-2014-00008
Device Sequence Number1
Product Code MEG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/08/2014,06/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2018
Device Catalogue NumberSG2-2038
Device Lot Number130716B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age10 MO
Event Location Hospital
Date Report to Manufacturer05/08/2014
Initial Date Manufacturer Received 08/15/2014
Initial Date FDA Received06/04/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ALKERMES MANUFACTRUED VIVITROL KIT
-
-