ST. JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Coiled (1098); Detachment Of Device Component (1104); Device Operates Differently Than Expected (2913)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/25/2014
Event Type  Injury  
Event Description
It was reported the pt was not receiving stimulation to her back, hips and legs.The pt underwent surgical intervention to replace the leads.During the procedure, it was revealed the leads were coiled together in a pocket and not down to the ipg.It was noted that the anchors were loose and not secured to the leads.Both leads were explanted but the physician was only able to implant one new lead.Strain relief loops were used and the anchors were sutured securely.Postoperatively the pt was reprogrammed and she was receiving effective stimulation.The leads will not be returned to the manufacturer for analysis.
 
Manufacturer Narrative
Sjm has limited info related to the pt's medical history and is unable to form an opinion as to the relevancy of the pt's history to the event reported.Sjm defers to the pt's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
tanja hoffman
6901 preston rd.
plano, TX 75024
9723098520
MDR Report Key3850291
MDR Text Key4410106
Report Number1627487-2014-26418
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 04/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/01/2015
Device Model Number3186
Device Lot Number4349932
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/25/2014
Initial Date FDA Received05/15/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANTED:; SCS IPG: MODEL 3716
Patient Age74 YR
-
-