Received one intact 14.5f x 24 cm hemo-flow catheter for evaluation.A functional examination revealed a hole in the venous side of the lumen at the lumen to hub junction that leaks when the device is flushed.The hole is only visible when the lumen is manipulated away from the hub.A review of the mfg records for the device involved in the event was not possible as the lot number of the device was not provided.We are unable to determine the cause or factors that may have contributed to this event.An implanted catheter is exposed to many variables for which the mfr has no influence or control.These variables include but are not limited to: pt physiology, care and maintenance of the catheter by healthcare providers and the pt, exposure to site care agents and ointments, locking solutions, and off label uses such as medication admin.The longer the catheter is implanted and functioning properly the less likely a catheter's failure can be contributed to a mfg defect.The device involved in this event was implanted for 10+ months.
|