• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO BROCKENBROUGH¿ NEEDLE; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO BROCKENBROUGH¿ NEEDLE; TROCAR Back to Search Results
Model Number EP003994S
Device Problems Break (1069); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2014
Event Type  malfunction  
Event Description
It was reported that when the physician was inserting the needle into a competitor transseptal sheath, a piece of plastic from the sheath "went out." no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
Evaluation summary: analysis was performed and no anomalies were found.
 
Manufacturer Narrative
Product event summary: the transseptal needle, ep003994s with lot number 207837995, was returned and analyzed.Analysis was performed and no anomalies were found.It is plausible that skiving occurred as there is a compatibility issue with the needle and a competitor sheath.The cause of the issue was not established.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH¿ NEEDLE
Type of Device
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key3859561
MDR Text Key4656883
Report Number9612164-2014-00591
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/18/2015
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number207837995
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LAMP 45 COMPETITOR TRANSSEPTAL SHEATH
-
-