MEDTRONIC MEXICO BROCKENBROUGH¿ NEEDLE; TROCAR Back to Search Results
Model Number EP003994S
Device Problems Break (1069); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/26/2014
Event Type  malfunction  
Event Description
It was reported that the physician experienced difficulties inserting the needle into a competitor transseptal sheath.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the physician experienced difficulties inserting the needle into a competitor transseptal sheath, and that a piece of plastic was produced which got out of the sheath.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the transseptal needle, ep003994s with lot number 207837990, was returned and analyzed.There were no anomalies found with the needle.However, it is plausible that skiving occurred as there is a compatibility issue with the needle and a competitor sheath.The cause of the issue was not established.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH¿ NEEDLE
Type of Device
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key3859639
MDR Text Key4686302
Report Number9612164-2014-00589
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 03/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/18/2015
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number207837990
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/18/2014
Initial Date FDA Received06/10/2014
Supplement Dates Manufacturer ReceivedNot provided
01/22/2020
02/28/2020
Supplement Dates FDA Received10/10/2014
01/23/2020
03/05/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LAMP 45 COMPETITOR SHEATH
-
-