• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO VIVA S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO VIVA S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBB1D1
Device Problem Component Missing (2306)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2014
Event Type  malfunction  
Event Description
It was reported that during an implant procedure, the setscrew for the right ventricular high power port was not in the grommet so the physician could not engage the setscrew with the wrench.The device was not used and another device was implanted.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary : the device was returned and analyzed with no anomalies found.The set screw was noted to be loose/detached.Concomitant products: 5076-52 lead, implanted (b)(6) 2004; 6947-65 lead, implanted (b)(6) 2004.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA S
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3859734
MDR Text Key4686765
Report Number3004209178-2014-10713
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/14/2015
Device Model NumberDTBB1D1
Device Catalogue NumberDTBB1D1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4193-88 LEAD
Patient Age00085 YR
-
-