MEDTRONIC, INC. ATTAIN VENOGRAM BALLOON CATHETER; CATHETER, FLOW DIRECTED Back to Search Results
Model Number 6215
Device Problem Deflation Problem (1149)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/01/2014
Event Type  Injury  
Event Description
It was reported that during the implant procedure the balloon catheter would not deflate while in the patient.The physician was unable to deflate the balloon by applying negative pressure with a syringe, and the catheter had to be removed using force.The implant was completed using a new balloon catheter.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
Product event summary: the catheter was returned, analyzed, and no anomalies were found.Visual summary analysis of the catheter indicated damage at implant.The analyst noted that the leg with the syringe connection is cut off from the remaining portion of the catheter.Balloon inflation/deflation could not be tested.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN VENOGRAM BALLOON CATHETER
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3862556
MDR Text Key4462190
Report Number2182208-2014-01685
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K012225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6215
Device Catalogue Number6215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/01/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-