• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 97714
Device Problems Failure to Deliver Energy (1211); Device Difficult to Program or Calibrate (1496); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Type  malfunction  
Event Description
The patient experienced a shocking or jolting sensation.The patient was reprogrammed this past wednesday.Since the reprogramming, the stimulation has been turning on/off by itself.The adaptive stim (as) was all reset to a different setting.The as did not recognized the new setting, has gone back to old setting, even after the 3 minute timer.When the patient was lying on her back, she decreased her stimulation to 0v, set her phone timer to 3 minutes, raised and bent her leg.After the 3 minutes, the device went up to 1.70v and it was displaying lying on her back.At that time the patient felt jolting from her stimulation.The ins was not recognizing the last setting she programmed with her patient programmer.The only position the ins displayed was lying on her back, no matter what position the patient was in.The ins pocket did not seem loose nor did it seem like the device was moving around.The company representative palpated the ins pocket and it did not seem loose to him.The physician programmer (8840) displayed the following: lying on back when in a sitting position, lying on back when lying on right side, lying on back when lying on back.Additional information has been requested.
 
Manufacturer Narrative
Concomitant medical products: product id: 39565-65, serial# (b)(4), implanted: (b)(6) 2014, product type: lead.Product id: 97740, serial# (b)(4), product type: programmer, patient.Product id: 97754, serial# (b)(4), product type: recharger.(b)(4).
 
Event Description
It was later reported that the patient¿s device was reoriented and the shocking sensation was eliminated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key3869701
MDR Text Key4512371
Report Number3004209178-2014-11479
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/14/2014
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/11/2014
Initial Date FDA Received06/12/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/18/2014
Date Device Manufactured07/20/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00040 YR
-
-