MEDTRONIC MED REL MEDTRONIC PUERTO RICO SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
|
Back to Search Results |
|
Model Number 97714 |
Device Problems
Failure to Deliver Energy (1211); Device Difficult to Program or Calibrate (1496); Inappropriate/Inadequate Shock/Stimulation (1574)
|
Patient Problem
Electric Shock (2554)
|
Event Type
malfunction
|
Event Description
|
The patient experienced a shocking or jolting sensation.The patient was reprogrammed this past wednesday.Since the reprogramming, the stimulation has been turning on/off by itself.The adaptive stim (as) was all reset to a different setting.The as did not recognized the new setting, has gone back to old setting, even after the 3 minute timer.When the patient was lying on her back, she decreased her stimulation to 0v, set her phone timer to 3 minutes, raised and bent her leg.After the 3 minutes, the device went up to 1.70v and it was displaying lying on her back.At that time the patient felt jolting from her stimulation.The ins was not recognizing the last setting she programmed with her patient programmer.The only position the ins displayed was lying on her back, no matter what position the patient was in.The ins pocket did not seem loose nor did it seem like the device was moving around.The company representative palpated the ins pocket and it did not seem loose to him.The physician programmer (8840) displayed the following: lying on back when in a sitting position, lying on back when lying on right side, lying on back when lying on back.Additional information has been requested.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: 39565-65, serial# (b)(4), implanted: (b)(6) 2014, product type: lead.Product id: 97740, serial# (b)(4), product type: programmer, patient.Product id: 97754, serial# (b)(4), product type: recharger.(b)(4).
|
|
Event Description
|
It was later reported that the patient¿s device was reoriented and the shocking sensation was eliminated.
|
|