Eval of the device was completed by the mfr.Visual inspection found that fittings were missing from the water trap and oxygen connector, an extra flow meter was attached to the side of the device and one of the screws is missing from the alarm block.To test the device a test fitting was placed on the oxygen connector.Testing of the mixer found that the proportioning valve would only read 21.2% oxygen and there was no flow output.No further testing could be completed.Upon turning off the gas flow, the 100-1000 ml/min float ball would stick to the top of the flow meter and then drop down.As further testing could not be completed, the mixer was further inspected.The lock nut on the proportioning valve and the check valve on the retainer were found to be loose.Further examination of the proportioning valve found that one of the o-rings appeared to be cut.In addition, the fio2 adjuster needle was turned out too far.Per the customer, the mixer had been overhauled in 02/2014 by a third party.The third party is not recognized by the mfr as having been trained to complete maintenance and/or repairs on the mfr's devices.The root cause for the failure of the mixer is tampering and/or attempted maintenance.The device was not calibrated properly during the overhaul in 02/2014.The root cause for the failure of the float ball in the flow meter is age and/or wear and tear.The flow meter is the original flow meter from time of the mfr and is 26 years old.
|