Medtronic received information indicating that during ablation, as the surgeon was repositioning this cardioblate gemini and removing the cables from the device, the connector on the top jaw became disconnected.The device was replaced and the procedure was completed with no adverse patient effects reported.The product was returned for analysis.
|
Upon receipt at medtronic's quality laboratory, visual inspection confirmed that the hooded shroud was missing.The hooded shroud was returned with the device.The sheath was observed under magnification, and there appeared to be adhesive residue on it.Medtronic's quality engineers could not identify the root cause of the event, as there was evidence of adhesive on the sheath of the device where the shroud was attached.These products are 100% inspected prior to release for distribution.There are no trends associated with this issue.(b)(4).
|