The facility is using olympus scopes and using the reusable olympus maj-855 auxilliary water tubing at bedside for flushing.The medivators clinical specialist was doing an inservice on the medivators dsd-201 & dsd-edge machines and discovered this facility is not properly disinfecting the maj-855 between patient procedures.The facility stated that they only put alcohol through it at the end of the day.The medivators clinical specialist informed them that the maj-855 needs to be reprocessed between every patient for there is potential for cross contamination.The steps of appropriate reprocessing of the maj-855 is written in the olympus ifu for this product as well as a medivators product bulletin.To date, there are no reports of patient injury or illness.This complaint will be monitored within medivators complaint system.
|