Model Number 97714
Device Problems High impedance (1291); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2014
Event Type  malfunction  
Event Description
It was reported that during a replacement procedure for normal battery depletion, impedance testing with a snaplid showed electrode 3 with high impedances.Rest of the results were within normal limits.When the lead was connected to the new device, the bottom port felt difficult to advance into.The top lead had been placed first.The lead had been taken out and wiped down several times.High impedances continued even when tested at 3v.The leads were swapped (bottom lead to top port/vice versa) and high impedance was seen on both leads.When both leads were taken out and placed on the snaplid again, electrode 10 was showing high impedance.The physician planned to switch out the stimulator.It was further noted that the bottom lead went in much easier in the second stimulator that was opened and placed.It was clarified that there were normal readings prior to attaching to the initial stimulator but not when attached.
 
Manufacturer Narrative
Concomitant products: product id 3778-60, serial # (b)(4), implanted: (b)(6) 2008, product type lead; product id 3778-60, serial # (b)(4), implanted: (b)(6) 2008, product type lead.Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.
 
Manufacturer Narrative
Final device analysis of the stimulator revealed no anomalies.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key3882041
MDR Text Key4510941
Report Number3004209178-2014-11777
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2015
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received06/19/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/05/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00032 YR
-
-