MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
|
Back to Search Results |
|
Model Number 97714 |
Device Problems
High impedance (1291); Connection Problem (2900)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/29/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during a replacement procedure for normal battery depletion, impedance testing with a snaplid showed electrode 3 with high impedances.Rest of the results were within normal limits.When the lead was connected to the new device, the bottom port felt difficult to advance into.The top lead had been placed first.The lead had been taken out and wiped down several times.High impedances continued even when tested at 3v.The leads were swapped (bottom lead to top port/vice versa) and high impedance was seen on both leads.When both leads were taken out and placed on the snaplid again, electrode 10 was showing high impedance.The physician planned to switch out the stimulator.It was further noted that the bottom lead went in much easier in the second stimulator that was opened and placed.It was clarified that there were normal readings prior to attaching to the initial stimulator but not when attached.
|
|
Manufacturer Narrative
|
Concomitant products: product id 3778-60, serial # (b)(4), implanted: (b)(6) 2008, product type lead; product id 3778-60, serial # (b)(4), implanted: (b)(6) 2008, product type lead.Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.
|
|
Manufacturer Narrative
|
Final device analysis of the stimulator revealed no anomalies.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|