• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C4479C
Device Problems Fluid/Blood Leak (1250); Human-Device Interface Problem (2949)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/25/2014
Event Type  malfunction  
Event Description
It was reported that a homechoice automated pd set with cassette had a small leak from the patient line.This was identified during dwell cycle one of four of peritoneal dialysis (pd) therapy while the patient was connected.The patient reported that they had opened the box of supplies with a razor knife and may have damaged the patient line in doing so.The technical service representative assisted the patient in ending therapy and advised the patient to restart therapy with new supplies.There was no patient injury or medical intervention reported.No additional information is available.
 
Manufacturer Narrative
(b)(4).As the cassette was not returned and the lot number is unknown, a device analysis cannot be completed.However, the patient reported that they made a mistake and punctured a hole in the patient line of the disposable cassette.Proper user instructions are addressed in ¿the homechoice and homechoice pro apd systems patient at-home guide¿ which is shipped with every homechoice device.The guide has instructions on how to properly handle the disposable set when opening the packaging in order to avoid damage.The guide also states that the user has to inspect the disposables cassette for damage and if damage is found, then the user is to ¿obtain a new disposable set¿ in order to avoid possible contamination of the fluid pathways.A review of the label for the product family will be conducted.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3884144
MDR Text Key4542203
Report Number1416980-2014-19470
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C4479C
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/25/2014
Initial Date FDA Received06/19/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HOMECHOICE
Patient Age37 YR
-
-