Model Number CD3231-40
Device Problems Inappropriate or Unexpected Reset (2959); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2012
Event Type  malfunction  
Event Description
It was reported that during dft the device went into bvvi as a result inappropriate shocks were delivered.Patient was defibrillated externally.The device was explanted.
 
Manufacturer Narrative
This historical complaint is being filed as part of a retrospective review of complaint files in response to a recent fda inspection.There is no change to the actual performance of the product and this report only represents an enhancement to the reporting criteria going forward.Evaluation description included.The reported filed event of bvvi was confirmed in the laboratory.The cause of the reset is due to external defibrillation that was used during dft testing at implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR, NIK
Manufacturer (Section D)
ST. JUDE MEDICAL INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
cary lawler
15900 valley view court
sylmar, CA 91342
8184932621
MDR Report Key3891066
MDR Text Key4678821
Report Number2938836-2014-03608
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2013
Device Model NumberCD3231-40
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/31/2012
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date01/23/2012
Device Age5 MO
Event Location Hospital
Initial Date Manufacturer Received 01/24/2012
Initial Date FDA Received01/13/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial