Model Number 20
Device Problems Device Inoperable (1663); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 05/23/2014
Event Type  malfunction  
Event Description
It was reported that the customer's device had logged multiple event codes.The device was also not powering on completely as it was power cycling itself during the boot up process.The device had recently been received back from repair from the third party service agent.There was no patient use associated with the reported failure.
 
Manufacturer Narrative
(b)(4).The third party service agent evaluated the device and verified the reported failure.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Manufacturer Narrative
The third party service agent replaced the system controller and user interface pcb assemblies and observed proper device operation thorugh functional and performance testing.The device was then returned to the customer for use.The device has not been returned to physio-control for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key3894530
MDR Text Key4647047
Report Number3015876-2014-00710
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age11 YR
Event Location Hospital
Initial Date Manufacturer Received 07/28/2014
Initial Date FDA Received06/25/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/01/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/21/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-