The account found the brake detent was broken.Per the hill-rom service manual the affinity three birthing bed and affinity four birthing beds require and effective maintenance program.We recommend that you perform semiannual preventive maintenance.Check the tires for cuts, wear, tread life, etc.Apply the brake, and check to ensure that the bed will not move.If the bed moves, inspect it for wear, and adjust if required.Apply the steering pedal and check the steering to ensure proper locking action when activated.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2010-2012.It is unknown if the facility performed any other preventative maintenance on this bed.The account replaced the brake detent to resolve the issue.Based on this information, no further action is required.
|