A pt was on single ventricle circulatory support utilizing hl20 and rotaflow.The rotaflow was in art mode and the inlet pressure was monitored on pressure 3 allowing for servo regulation.The clinician observed that the rotaflow/hl20 responded to an inlet pressure in the servo regulation range with a soft alarm and slowing rpms.The rpms continued to ramp down to the point that they reached zero.This situation was interpreted by the clinician as hypovolemia or cannulae misplacement.After attempts to correct the situation, the clinician chose to provide manual circulatory support.It was determined that a clot had developed in the pressure line at the point where the pressure tubing connects to the circuit.It is believed that the clot "fixed" the pressure at a point within the servo regulation range.Despite the release or reduction in pressure within the circuit, it was not reflected in the pressure line, and therefore, did not transmit to the pressure transducer and hl20 controller.Therefore the hl20 continued to interpret a pressure requiring servo regulation intervention.No adverse consequences were reported for this event.
|
The device was not returned and no pictures or tubing pack lot info was provided.A root cause could not be identified or the complaint confirmed as no device was available for testing.Additionally, since no pack lot info was provided, device history records could not be reviewed.If additional info is received or the product is returned, a supplemental report will be submitted.(b)(4).
|