GUIDANT CRM CLONMEL IRELAND AUTOGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G179
Device Problems Fitting Problem (2183); Impedance Problem (2950)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/21/2014
Event Type  malfunction  
Event Description
Boston scientific received information that during device implant, abnormal impedance measurements were exhibited following right and left ventricular lead-device connection.The physician disconnect and cleaned the terminal pins, prior to reinserting.The physician reported that during reconnection, additional force was used to push the terminal pins harder into the device header.To date, the system remains implanted and in service with no further reported complications.
 
Manufacturer Narrative
(b)(4).As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key3904517
MDR Text Key4517915
Report Number2124215-2014-11792
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/09/2015
Device Model NumberG179
Other Device ID NumberAUTOGEN X4 CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 04/15/2014
Initial Date FDA Received07/01/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/28/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0692; G179; 4674
-
-