(b)(4).Lockout premature, damaged latch post, jaw misaligned.The analysis results found that the mcm30 device was returned in good visual condition.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled, fed, and formed the remaining 1 clip as intended.The device was found to be fully functional and conforming to our manufacturing specifications.In addition, one malformed clip was received inside a sample bag.No conclusion could be reached as to what may have caused the reported incident or the received malformed clip.The batch record was reviewed and no anomalies were noted during the manufacturing process.
|