U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA FX OXY W/HR & ART FILTER; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA FX OXY W/HR & ART FILTER; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 1CX*FX25RE
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 02/17/2014
Event Type  malfunction  
Event Description
The user facility reported to terumo cardiovascular systems that prior to cardiopulmonary bypass, during setup, the sampling manifold that connects to the arterial outlet disconnected from the white luer connection.No pt involvement as this occurred during setup.The product was not used.The surgery was completed successfully.
 
Manufacturer Narrative
Terumo has received the actual device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is completed and more info becomes available.For this reason, (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FX OXY W/HR & ART FILTER
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, shizuoka, prefecture 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION
ashitaka, 150
Manufacturer Contact
eileen dorsey, manager
125 blue ball road
elkton, MD 21921
8002623304
MDR Report Key3908687
MDR Text Key15365660
Report Number9681834-2014-00064
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130520
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 02/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2016
Device Model Number1CX*FX25RE
Device Lot Number130510
Other Device ID Number(01) 00699753450462
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Device Age9 MO
Event Location Hospital
Date Manufacturer Received02/18/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-