STRYKER INSTRUMENTS-KALAMAZOO LOANER CORE IMPACTION DRILL; DRIVER, WIRE, AND BONE DRILL, MANUAL
|
Back to Search Results |
|
Catalog Number 5400300000L |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/11/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during a procedure at the user facility the loaner core impaction drill was overheating.The procedure was completed successfully using back-up equipment with no delay, no medical intervention, and no adverse consequences reported.
|
|
Manufacturer Narrative
|
The failure for heat was confirmed through functional evaluation, disassembly, and visual inspection.Through visual inspection the technician confirmed the bearings were corroded.The device was repaired and placed back into stryker stock.
|
|