Model Number 3186
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913); Impedance Problem (2950)
Patient Problem Inadequate Pain Relief (2388)
Event Date 05/05/2014
Event Type  Injury  
Event Description
Device i of 2.Reference mfr report #: 1627487-2014-08045.It was reported the pt was experiencing intermittent stimulation beginning on (b)(6) 2014.An sjm rep met with the pt and confirmed invalid scs system impedances.The rep was able to reprogram the pt to provide stimulation in the desired areas but the stimulation was abrupt when turning-up the amplitude.Pt met with his physician on (b)(6) 2014 and had a ct scan.Pt is going to f/u with his physician to evaluate results of ct and determine the next course of action to address the issue.
 
Manufacturer Narrative
Sjm has limited info related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
logan farmer
6901 preston rd.
plano, TX 75024
9725269611
MDR Report Key3919138
MDR Text Key19298259
Report Number1627487-2014-08044
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2010
Device Model Number3186
Device Lot Number115539
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/23/2014
Initial Date FDA Received06/16/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS IPG: MODEL 3716
Patient Outcome(s) Other;
Patient Age78 YR
-
-