• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST JUDE MEDICAL- NEUROMODULATION EON MINI; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST JUDE MEDICAL- NEUROMODULATION EON MINI; SCS IPG Back to Search Results
Model Number 3788
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem Inadequate Pain Relief (2388)
Event Date 05/19/2014
Event Type  Injury  
Event Description
Device 2 of 2: ref mfr report: 1627487-2014-02418.The pt received two scs leads for the same lot number.It was reported the pt is not receiving any stimulation therapy from his scs system.The pt stated his stimulation had been intermittent.A diagnostics of the pt's system revealed all the contacts are invalid for both leads.Additional info received on (b)(6) 2014 identified the pt's scs lead were explanted and replaced.It was observed the leads appeared fractured with fluid inside the casing.In addition, it was uncertain if any fluid intruded in the pt's ipg, therefore, the physician decided to replace the pt's ipg.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI
Type of Device
SCS IPG
Manufacturer (Section D)
ST JUDE MEDICAL- NEUROMODULATION
plano TX
Manufacturer Contact
arnie ochoa
6901 preston rd.
plano, TX 75024
9723098090
MDR Report Key3926149
MDR Text Key4582988
Report Number1627487-2014-02419
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/01/2015
Device Model Number3788
Device Lot Number4271699
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age87 YR
-
-