• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-ASD-018
Device Problems Migration or Expulsion of Device (1395); Infusion or Flow Problem (2964)
Patient Problem Embolism (1829)
Event Date 06/18/2014
Event Type  Injury  
Event Description
The patient's defect measured 15mm based on sizing balloon and stop-flow technique and an 18mm amplatzer septal occluder (aso) was implanted after verifying its stability in the defect.A trans-thoracic echo verified there was no residual flow across the defect.A follow-up echo the next day, however, showed the aso had embolized into the left ventricular outflow tract touching the anterior mitral valve leaflet.The patient was referred for surgery and the aso was explanted.
 
Manufacturer Narrative
The aso was not returned to sjm for analysis.The device history record for this product was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with sjm specifications and procedures.The results of this investigation are inconclusive because the product was not returned for analysis; however, review of the device history record confirmed the product met manufacturing requirements prior to shipment.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key3926962
MDR Text Key19764979
Report Number2135147-2014-00068
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeEG
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date07/31/2018
Device Model Number9-ASD-018
Device Lot Number4244266
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/19/2014
Initial Date FDA Received07/10/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/26/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age7 YR
Patient Weight21
-
-