Catalog Number 5400300000 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/17/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during a procedure at the user facility, the core impaction drill was overheating.The procedure was completed successfully using back-up equipment with no delay; no adverse consequences or medical intervention were reported.
|
|
Manufacturer Narrative
|
Failure analysis is in progress; additional information will be submitted once the quality investigation is completed.
|
|
Event Description
|
It was reported that during a procedure at the user facility, the core impaction drill was overheating.The procedure was completed successfully using back-up equipment with no delay; no adverse consequences or medical intervention were reported.
|
|
Manufacturer Narrative
|
The reported event was duplicated.Upon disassembly for visual inspection, the service technician observed bearing corrosion.
|
|