Catalog Number 5400300000 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/19/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during a third molar case at the user facility the core impaction drill was overheating.The procedure was completed successfully using back-up equipment without a clinically significant delay; no medical intervention and no adverse consequences were reported.
|
|
Manufacturer Narrative
|
Failure analysis is in progress; additional information will be submitted once the quality investigation is completed.
|
|
Event Description
|
It was reported that during a third molar case at the user facility, the core impaction drill was overheating.The procedure was completed successfully using back-up equipment without a clinically significant delay; no medical intervention and no adverse consequences were reported.
|
|
Manufacturer Narrative
|
The technician was able to duplicate the reported event and noted that the driveshaft bearing was damaged.
|
|